LEGEND
open squares: reversal obtaining low-pressure phase
closed squares: reversal obtaining high-pressure phase
diamonds: electrical resistivity change accompanying Bi transition
This graph shows the
pressure calibration for the 14/8 (ram force vs. the pressure of "fixed"
transitions). The Walker module is used for achieving pressures up
to 25 GPa. In this device, six anvil "wedges" press on 8 cubes which in
turn press on an octahedra press medium. The press is calibrated
using a variety of sample assemblies, using both castable ceramic and
semi-sintered ceramic pressure media. This configuration is used
at
Dr. Paul Asimow's lab at CalTech. The 14/8 ass'y is
configured as follows:
>>14 mm edge length octahedron made of semi-sintered MgO plus 5 wt% Cr2O3
>>Toshiba grade-F WC cubes with a truncated edge length of 8 mm
>>Stepped LaCrO3 heater
>>Pyrophyllite gaskets with a cross section of 2.8 x 4.5 mm
Rockland Research Corporation
16 Tompkins Avenue
West Nyack, NY 10994
Phone: (845) 353-4686
Fax: (845) 358-9119
web design by
(Pressure chart courtesy of CalTech Professors Paul Asimow and Jed Mosenfelder)